The temperature display area displays the patient temperature. Spot Vital Signs displays the current temperature for two minutes after the probe is placed back in the holder. The display then goes blank unless another measurement is active.
Page Monitor Mode Directions for Use Monitor Mode Monitor Mode displays the temperature of the probe as long as the probe remains in place at the measurement site and remains within the operating patient temperature range. Page Error Indications And Interpretation Error Indications and Interpretation The following table of conditions and error codes provides a quick reference of the descriptions and probable causes of error codes.
To clear the error code: Power the Spot Vital Signs off, wait five seconds, and power on. If the error code reappears then power the Spot Vital Signs off and disconnect the battery for five minutes. Loss of tissue contact Ensure proper probe positioning. The Welch Allyn Spot Vital Signs was developed using the American Heart Association recommendations, which state that phase 5 be used unless sound continues to 0 mmHg, in which case the change in the quality of sound phase 4 is to be used.
Page 48 Verify the sensor cable is correctly plugged into device. Verify you are using the correct sensor. Table Device Does Not Turn On Page 49 Welch Allyn Technical Support. Blood pressure cuff applied Re-apply blood pressure cuff. Make sure Welch Allyn label is facing away from arm. Page Specifications Welch Allyn defines a pediatric patient as 29 days or more of age. Welch Allyn defines neonates as children 28 days or less of age, born at term 37 weeks gestation or more , otherwise up to 44 gestational weeks.
Pulse oximeter readings were compared to SaO values of drawn blood samples measured by hemoximetry. The customer or user of the Series Spot Vital Signs should assure that it is used in such an environment. Page Maintenance Maintenance Welch Allyn will make available,upon request, circuit diagrams and other information which will assist appropriately qualified technical personnel in repair of this device.
Using the incorrect battery will cause damage to the Spot Vital Signs and void the warranty. As necessary, replace the internal battery after heavy use or the battery no longer charges. Slide the new battery completely into the compartment. Lay the connector on the battery. The relief pocket in the battery door purposely provides sufficient clearance for the battery connector. Relief Pocket Battery Battery Connector 5. Replace the battery door and tighten each of the 4 screws.
If you are advised to return a product to Welch Allyn for repair or routine maintenance, schedule the repair with the service center nearest you. Cuff and Bag Combination Catalog Description Catalog Description Cuff and bladder, adult, one tube Cuff and bladder, thigh, one tube Cuff and bladder, large adult, one tube Cuff and bladder, child, one tube Table 2.
DEC-8 extension cable, 8 ft. Page Warranty Warranty Spot Welch Allyn warrants Spot Vital Signs, when new, to be free of defects in material and workmanship and to perform in accordance with manufacturer's specifications for a period of two years from the date of purchase from Welch Allyn or its authorized distributors or agents. Page 74 Reorder No. This manual is also suitable for: tb 42mob 42ntb 42nob 42mtb. Print page 1 Print document 74 pages. Rename the bookmark. Delete bookmark?
Cancel Delete. Forgot password? Enter your email address and check your inbox. Please check your email for further instructions. Enter a new password.
Welch Allyn. Spot Vital Signs B. You can only view or download manuals with. Sign Up and get 5 for free. Upload your files to the site. You get 1 for each file you add. Get 1 for every time someone downloads your manual. Buy as many as you need. View and download manuals available only for. Register and get 5 for free.